Working Advice
Marking objects
If required, detected objects can be marked. Perform a measurement as usual.
Once you have found the boundaries or the centre of an object, mark the sought location at the top marking aid (7) and the side marking aid (6). Connect the points with a vertical and horizontal line. The boundary or the centre of the object is located where the lines intersect.
Recalibration
Manually recalibrate the measuring tool if the light-up ring (5) continues to light up yellow or red even though there is no metal near the measuring tool.
- Switch on the measuring tool using the on/off switch (2).
- Remove a battery from the measuring tool whilst it is switched on.
- Switch off the measuring tool using the on/off switch (2) after the battery has been removed.
- Reinsert the battery into the measuring tool. Ensure that the polarity is correct.
- Remove all objects near the measuring tool (including watches or metal rings) and hold the measuring tool in the air.
- Switch the measuring tool on and off again within 3 s using the on/off switch (2). The light-up ring (5) will flash red slowly for the 3 s to indicate readiness for calibration.
- Switch the measuring tool back on within 0.5 s. The calibration will start and take roughly 6 s. During calibration, the light-up ring (5) flashes green quickly. Once the light-up ring (5) lights up green continuously, the calibration is complete and the measuring tool is ready to use.
Note: If the sequence of switching off and switching on again is not observed, calibration will not take place. The light-up ring (5) continues to light up yellow or red even though there is no metal nearby. In this case, repeat the calibration in the exact same way.